DTIC ADA084090: Laboratory Evaluation of the USDA pdf

DTIC ADA084090: Laboratory Evaluation of the USDA_bookcover

DTIC ADA084090: Laboratory Evaluation of the USDA

More Book Details

Description of the Book:

Tests conducted in the laboratory determined that the insect detector (second prototype) could distinguish between a sample with three live insects per pound and an uninfested sample of stored food products. The detector could not differentiate between no infestation and one insect in 5-pound packaged commodities or a 1-pound sample of whole grain. The length of time necessary to test a sample and extrinsic interferences from various sources made the first and second prototypes of the hidden insect detector operationally unsatisfactory. (Author

  • Creator/s: Defense Technical Information Center
  • Date: 2/1/1980
  • Year: 1980
  • Book Topics/Themes: DTIC Archive, Anderson, Leroy M, ARMY MEDICAL BIOENGINEERING RESEARCH AND DEVELOPMENT LAB FORT DETRICK MD, *DETECTORS, *BIOLOGICAL DETECTION, *INSECTS, TEST AND EVALUATION, SAMPLING, LOW LEVEL, PACKAGING, STORAGE, FOOD